Makmal Biokeselamatan PCR P2

Makmal Biokeselamatan PCR P2

It is mainly used for primary health services, diagnosis and research. The hazard level of the experimental subjects is II (medium individual hazard, limited group hazard), which is specifically defined as "can cause disease in humans or animals, but generally affect health workers and groups , Livestock or the environment will not cause serious harm to pathogens. Laboratory infections do not cause serious diseases, effective treatment and preventive measures are available, and the risk of transmission is limited."
Hantar pertanyaan
Berbual sekarang

Maksud makmal P2 Makmal Biokeselamatan PCR P2:


Merujuk kepada klasifikasi tahap keselamatan makmal biologi. Di antara semua jenis makmal, makmal P2 ialah makmal biokeselamatan yang paling banyak digunakan. Tahap makmal biologi ialah P1, P2, P3, dan P4. Mengikut sifat makmal untuk memilih tahap yang sepadan. Pertubuhan Kesihatan Sedunia (WHO) membahagikan mikroorganisma berjangkit kepada 4 kategori mengikut patogenik dan risiko jangkitan; mengikut peralatan dan keadaan teknikal, makmal biologi juga dibahagikan kepada 4 peringkat (biasanya dipanggil P1, P2, P3), makmal P4), tahap 1 adalah yang paling rendah, tahap 4 adalah yang tertinggi. Makmal P1 secara amnya sesuai untuk mikroorganisma yang tidak mempunyai kesan patogenik pada orang dewasa yang sihat; makmal P2 sesuai untuk mikroorganisma yang mempunyai potensi bahaya sederhana kepada manusia dan alam sekitar; makmal P3 sesuai untuk jangkitan serius atau serius terutamanya melalui laluan pernafasan. Mikroorganisma patogen atau toksin mereka penyakit maut; Makmal P4 sesuai untuk mikroorganisma patogen atau toksinnya yang sangat berbahaya kepada tubuh manusia, dihantar melalui aerosol atau tidak diketahui, dan tiada vaksin atau kaedah rawatan yang berkesan.

Penggunaan percubaan P2:


It is mainly used for primary health services, diagnosis and research. The hazard level of the experimental subjects is II (medium individual hazard, limited group hazard), which is specifically defined as "can cause disease in humans or animals, but generally affect health workers and groups , Livestock or the environment will not cause serious harm to pathogens. Laboratory infections do not cause serious diseases, effective treatment and preventive measures are available, and the risk of transmission is limited."


The national standard "Technical Specifications for Building Biosafety Laboratories" stipulates the following technical indicators (static) for the BLS-2 laboratory:

1. Kebersihan: tiada keperluan

2. Perbezaan tekanan bilik bersebelahan dengan arah luar: -20Pa

3. Suhu: 18-27 darjah

4. Kelembapan relatif: 30-70 peratus

5. Pencahayaan minimum: 300LX


Penunjuk di atas adalah keperluan minimum untuk menubuhkan makmal P2. Pembina boleh meningkatkan penunjuk yang sepadan dengan sewajarnya mengikut keperluan sebenar, seperti kebersihan, perbezaan tekanan dan keperluan lain.